Evidence-Generated Sockets for Transtibial Prosthetic Limbs Compared With Conventional Computer-Aided Designs: A Multiple-Methods Study From the Patient’s Perspective

Evidence-Generated Sockets for Transtibial Prosthetic Limbs Compared With Conventional Computer-Aided Designs: A Multiple-Methods Study From the Patient’s Perspective

Evidence-Generated Sockets for Transtibial Prosthetic Limbs Compared With Conventional Computer-Aided Designs: A Multiple-Methods Study From the Patient’s Perspective

1Mechanical Engineering Department, University of Southampton, Mailpoint M7, University Road, Highfield, Southampton, United Kingdom

2Radii Devices Ltd, Bristol, United Kingdom

3Portsmouth Hospitals NHS Trust, Portsmouth, United Kingdom

4Opcare Ltd, Abingdon, United Kingdom

5St George's Hospital, London, United Kingdom

6North Bristol NHS Trust, Bristol, United Kingdom

7Oxford University Hospitals NHS Trust, Oxford, United Kingdom

8Nuffield Department of Orthopaedics and Musculoskeletal Sciences, University of Oxford, Oxford, United Kingdom

Corresponding Author:

Alex Dickinson, PhD


Background: Personalized prosthetic socket design depends upon highly skilled prosthetists. They aim to balance functional human-prosthesis coupling with safe, comfortable load transmission from the prosthesis to the skeleton, through vulnerable skin and soft tissues. Both traditional plaster and computer-aided design and manufacturing (CAD/CAM) methods are iterative, and sharing knowledge is difficult. Evidence-generated (EG) sockets derived from past computer-aided socket design (CASD) records could provide a personalized starting point for limb fitting, potentially reducing time spent on basic design and enabling prosthetists to focus on more highly-skilled customization.

Objective: This study aimed to assess the comfort of EG sockets, generated from past CASD records.

Methods: A crossover trial compared EG sockets, derived from 163 previous transtibial devices, with conventional clinician-led CAD/CAM sockets. Noninferiority was assessed for the socket comfort score (SCS) outcome measure, and semistructured interviews provided in-depth user analysis. The setting was 3 UK National Health Service clinics, with 17 participants with 19 transtibial amputations.

Results: EG sockets had no statistically significant difference in comfort compared with clinician-led control sockets (median SCS 8.6 for EG sockets and 8.8 for CAD/CAM controls; P=.43, effect size=0.05), but a lower variability in SCS across the group (95% CIs 8.0‐9.0 for EG and 7.5‐9.5 for CAD/CAM devices, respectively). Analysis of interviews revealed themes around fitting session experiences, similarities, and differences between the EG and CAD/CAM control sockets, and residual limb factors impacting perceptions of socket comfort. These provided insights into the participants’ experience of the study and the value of expert prosthetist input in socket design.

Conclusions: EG sockets demonstrated noninferiority to conventional clinical CASD practice in terms of socket comfort. Both quantitative and qualitative results indicated how clinician input remains essential and is valued by prosthesis users. Work is underway to incorporate the EG sockets into CASD software such that they can act as a digital starting point for modification by expert clinicians at fitting, potentially reducing time spent on basic design, enabling prosthetists to focus on more highly-skilled customization and co-design with their patients.

Trial Registration: ClinicalTrials.gov NCT06597266; https://clinicaltrials.gov/study/NCT06597266

JMIR Rehabil Assist Technol 2025;12:e69962

doi:10.2196/69962

Keywords



The lower limb prosthetics community has worked since the 1980s toward computer-aided design and manufacturing (CAD/CAM) technologies to support prosthetic socket design and fabrication workflows [1]. Within CAD/CAM, computer-aided socket design (CASD) is defined as the strategic modification of a 3D digital representation of the residual limb shape in software by an expert human prosthetist, producing a rectified socket shape design. The “CAM” typically refers to fabricating the socket by producing a corresponding rigid foam mold carved using a computer numerical-controlled robotic carver, followed by draping or lamination. CAD/CAM have been described as offering significant potential benefits over conventional plaster of Paris approaches [2], including reduced manual work time and exposure to plaster, an occupational health risk. This could allow clinicians to spend a greater proportion of their time in direct patient interactions, which might enhance patient engagement and facilitate shared decision-making [2]. Early CAD/CAM results were comparable with traditionally produced sockets but took longer and needed more adjustments [1-4]. However, over the learning curve, today clinicians using CAD/CAM can achieve clinical results that are comparable while saving time and delivering a better quality of life outcome [5-8].

These technologies were also proposed to offer opportunities for evidence-based decision support in socket design [1]. CASD generates a perpetual, quantitative design record, whereas the plaster design is destroyed upon socket fabrication. A digital record has value for education, peer support on complex cases, and if implemented for clinical decision support, further time savings to focus on the most highly skilled and value-added parts of socket personalization [3,9]. However, despite proposals since the 1980s to augment CAD/CAM by providing the ability to refer to previous design records [10], it appears that the full potential of the benefits offered through these rich datasets to improve the prosthetic rehabilitation process has not been fully exploited.

In particular, data might be used for generating socket design recommendations, which we define as evidence-generated (EG) sockets. Since Dean and Saunders in 1985 [11], clinical innovators have considered an alternative to performing CASD in a manner analogous to manual work with plaster, whereby a user could apply averaged rectifications from previous designs as an “overlay,” or select from a database of different size and shape “primitive” or “template” sockets, and scale and adjust them to fit to a newly presenting patient’s residual limb shape. This has been demonstrated for transtibial [3,9,12-14] and transfemoral socket designs [15,16], and in orthotics for scoliosis braces [17,18]. Recent publications present alternative methods of informing the socket design process by data, including fuzzy logic or inference to map linguistic descriptions of socket design approaches and patient descriptors, for application to new people [19,20], simulation of load-bearing and optimization for automating transradial socket design [21,22]. However, these studies have only been demonstrated in research settings and have not yet been clinically applied. Anecdotally, today, most CASD software packages in clinical use define templates as structured design workflows or sequences to guide users in applying their own choice of rectifications and gross design features.

The application of such methods would benefit from a systematic study of rectification design practice, although this is limited in the scientific literature [23] since the “Automated Fabrication of Mobility Aids” project [12]. Recently, researchers have begun to leverage high-resolution 3D scans and CASD data to investigate rectification sizes in transtibial and transradial design [24-26], and most recently, probabilistic methods have been used to derive insights into transtibial socket design strategy through the associations between design features [27]. Building upon those insights, this study aims to evaluate an EG socket concept developed by Radii Devices Ltd and the University of Southampton, with UK service provider Opcare Ltd providing data and expert clinical design insight. The objective was to compare the EG socket design approach with clinician-led CASD, using socket comfort outcome measures and capturing the patients’ experiences through qualitative interviews.


This study is reported using the STROBE (Strengthening the Reporting of Observational studies in Epidemiology) cross-sectional reporting guidelines [28].

Patient and Public Involvement and Engagement

The study research question was informed by patient and public involvement and engagement (PPIE) group discussions [29], which highlighted how socket comfort is paramount but difficult to achieve, and that delays between assessment and device provision can impair fitting. PPIE contributors also expressed support for sharing knowledge between prosthetics centers to enable service improvement. Collaboration with the Alex Lewis Trust during the study development reinforced this and provided review and feedback of the study design, recruitment posters, participant information sheets, and consent forms. PPIE contributors are also actively involved in disseminating the study findings to patients.

Study Design

The study used a single-blind, crossover design to assess comfort at socket fitting, followed by a qualitative study of semistructured interviews. An EG transtibial draped thermoplastic check socket was compared with a control CAD/CAM socket, which was designed by a prosthetist using Tracer CASD software (Ohio Willowwood Co) from the same residual limb 3D scan (Figure 1). Participants used the same interface as intended in the definitive device (ie, liners or socks). Quantitative and qualitative data were collected and analyzed independently and interpreted together. This study design was selected to capture the patient’s comparative experience of socket fitting from quantitative and qualitative perspectives, developed from a foundational CAD/CAM versus traditional socket comparison study [5] and with recent precedent in clinical assessment of adjustable sockets [30]. This study design was chosen because, first, new digital design and fabrication technologies are often considered in small scale or low technology readiness level trials in the scientific literature but are often not tested in power-sized, blinded, or controlled trials, with consideration of qualitative service user experience alongside quantitative outcome measures [23]; and second, it is important to understand the patient’s perceived experience. This may relate to socket fit and comfort, for which detailed, open descriptions offer more nuanced insights than a simple socket comfort score (SCS) [31]. It may also relate to the user’s perspective of device design and fabrication, to promote shared decision-making [32] and its potential benefits regarding patient engagement in and understanding of care.

As such, the work aligns with recently published perspectives on ethical considerations for development and clinical translation of prosthetic technologies [33].

Figure 1. For an exemplar participant: (A) the residual limb scan over a sock; (B) the prosthetist’s socket rectification design; (C) the evidence-generated design before adding brimline; and (D) and (E) the corresponding fitted check sockets. In rectification design maps, the color red represents a carve or press-fit between socket and limb, white represents exact fit, and blue represents buildup or limb-socket gap.

Ethical Considerations

Ethical approval was granted by institutional (ERGO 76033.A3) and UK national review boards (IRAS 313408 and HRAREC 22/YH/0215). Inclusion criteria included people aged 18 years and older, with transtibial amputation, deemed ready for a new prosthetic socket by their prosthetist in their usual clinical pathway, and willing and able to tolerate trialing 2 sockets at a fitting session. Convenience sampling was used to identify participants at 3 UK prosthetic rehabilitation centers. All participants provided written, informed consent (Multimedia Appendix 1); and the presented data are selected to avoid identification. Participants were not compensated.

Evidence-Generated Socket Design Method

A socket generation method was applied, based upon evidence of expert practice, building upon previous work [27]. To review briefly, a dataset of 163 transtibial residual limb 3D scans and corresponding socket designs was accessed in .aop format (version AAOP1), including labeled landmarks, exported from Omega software (Ohio WillowWood Co) with cylindrical sampling at a maximum 3° spacing on a minimum of 90 slices. These were produced at a single, large UK physical enablement center by 4 prosthetists, all with a minimum BSc qualification, of whom 2 had more than 20 years of experience, one had 5‐10 years, and the other was a first-year graduate. The corresponding patient characteristics, such as gender, age, time since amputation, and reason for amputation, were also collated. A Statistical Shape Model was generated to describe the residual limb shape and size with a minimal number of dimensions, corresponding to principal “modes” of variation. The training datasets were aligned with the assistance of the landmarks, registered, and a principal component analysis was conducted using the full surface shapes of the residual limb scans. Socket design features of local rectifications (patella tendon bar carve, paratibial carves, fibula head build, distal end build, distal tibia build, anterior tibia build, and supracondylar carves) and gross volume change were extracted from the dataset. Bayesian inference was applied to analyze the probabilistic association between patient characteristics and principal modes of limb shape variation (inputs), and the extracted design features for the 163 limb-socket pairs. Finally, the same inputs were obtained for the study participants, and check sockets were created by automatically applying modifications to their landmarked residual limb 3D scans, which were predicted using the statistical model. These were applied to the limb shape scan and saved as a .stl mesh file (output).

Quantitative Study

The EG socket design was generated automatically except for the brim line, which the fabrication technician was requested to apply in the same location as the control, prosthetist-designed socket. The prosthetist then worked through their standard fitting and assessment procedure for both sockets (Multimedia Appendix 2), including an SCS. At the fitting appointment, the prosthetist chose which socket was trialed first, and the patient participant was blinded to which socket was prosthetist-designed, CAD/CAM, or EG. At the end of the session, the patient participant and their clinician were allowed free choice over which check socket design was to be used for the definitive prosthesis, on the basis of their mutual agreement without any intervention from the researchers.

A power calculation indicated a sample size of 19 was required to test noninferiority in a crossover trial (power=0.9, significance level=0.05, mean difference=0, noninferiority limit=1.21 [34], and population SCS SD 1.2 [24]). A Shapiro-Wilk test indicated the control SCS data were not normally distributed (P=.01), so nonparametric descriptive statistics were calculated with CIs using the bootstrap method [35], and the paired Wilcoxon Signed-Rank test was used to assess the statistical significance of the difference between the sockets.

Qualitative Study

Following the quantitative study, 2 semistructured interviews were carried out to capture participants’ views and experiences of in-clinic fitting of 2 sockets designed in different ways, and general usability of their new prosthetic socket. The first was immediately after completion of the socket fitting, and the second was 1 month afterwards (Multimedia Appendix 3). Interviews were audio-recorded, anonymized, and transcribed verbatim by a professional transcribing company. A member of the research team (FM) checked a sample of transcripts against the audio recordings for accuracy. The transcripts were analyzed using thematic analysis, which provided a flexible approach to ascertain a clear understanding of the comparability of socket comfort and views of the processes [36]. The thematic analysis approach consisted of (1) familiarization with the data, (2) coding by hand and using NVivo software (Lumivero), (3) generating initial themes, and (4) reviewing and finalizing the final themes to capture consistent patterns within the data and key meaning relevant to the research questions [36]. One team member (MD-H) led all stages of the analysis; 2 other members (FM and JB) coded different sections of the data and together agreed on the final themes, providing verification and allowing for a range of interpretations.


Recruitment

In total, 17 recruited participants with 19 residual limbs completed the socket assessment study between March and November 2023 (Table 1 and Figure 2). Furthermore, 6 prosthetists designed control sockets for between 1 and 9 participants. Approval was initially granted to compare transparent check sockets, but due to slow recruitment, a study protocol amendment was approved to compare the EG check socket to a definitive CAD/CAM prosthetist-designed socket, reflecting more common practice in the participating clinics.

In addition, 2 participants exited before completing the postassessment interview, and a further 3 were not contactable for the 1-month follow-up interview.

Table 1. Description of participants.
CharacteristicStatistical value
Side (n=19), n
Left8
Right11
Sex (n=17), n
Women1
Men16
Age (years; n=17), median (IQR)66 (51‐73)
Time since amputation (years; n=19), median (IQR)4.0 (2.5‐8.9)
Reason for amputation (n=19), n
Infection or diabetic foot6
Dysvascularity4
Trauma3
Ischemia or CLTIa3
Sepsis1
Peripheral Neuropathy1
Chronic osteomyelitis secondary to trauma1
Reasons for new socket (n=19), n
Socket too large or limb shrinkage12
Requested new suspension method3
Socket too small or tight2
Alignment incorrect1
N/Rb3
Participant activity level (n=17)
A1: In-house walking or transfers only1
A2: Walking on flat ground only6
A3: Normal everyday walking8
A4: Additional high impact or energy activities0
N/R2

aCLTI: critical limb-threatening ischemia; activity level as assessed by the participants’ normal clinical team.

bN/R: not reported.

Figure 2. Study design and recruitment diagram. CAD/CAM: computer-aided design and manufacturing.

Quantitative Study Findings

The unmodified EG sockets had no statistically significant difference in SCS compared with prosthetist-designed CAD/CAM control sockets (median SCS 8.6 for EG sockets and 8.8 for CAD/CAM controls; P=.43 and effect size=0.05 in the paired significance test, Figure 3). Lower variability in SCS was observed across the study group for the EG sockets than the control sockets (95% CIs 8.0‐9.0 and 7.5‐9.5, respectively). Of the 19 EG sockets, 9 were given a comfort score within the noninferiority limit (standard error of measurement 1.21 points [34]) of the control. In total, 6 EG sockets were rated as more comfortable than the CAD/CAM control, by between 2.0 and 3.5 points; 4 EG sockets demonstrated lower comfort than the CAD/CAM control, by 1.5 to 2.5 points.

Figure 3. Noninferiority crossover study results comparing socket comfort score between the prosthetist control and evidence-generated sockets, for 19 fittings across 17 participants, presented as a paired comparison chart. Color coding denotes higher (green), the same (blue), or lower (purple) socket comfort score for evidence-generated than prosthetist control devices, within the noninferiority threshold of 1.21 points.

Qualitative Study Findings

Due to the significant impact and diverse experiences related to socket comfort, thematic analysis identified several broader themes that were beyond this paper’s objective. Focus is therefore limited to specific themes that offer insights into the participants’ views and experiences relating to the sockets produced by the 2 design approaches (Table 2).

Table 2. Themes identified in semistructured interviews relating to comparison between comfort of the fitted sockets and reflections on the socket fitting process.
ThemeDescription
Theme 1Experiences of the comparative fitting session
Theme 2Similarity between the compared sockets
Theme 3Differences between the compared sockets
Theme 4Residual limb factors impacting perceptions of socket comfort
Theme 1: Experiences of the Comparative Fitting Session

Both devices were fabricated before the fitting session, and when discussing the session, participants reflected on several aspects that they felt impacted their experiences. Several participants described feeling that it went “well,” “fine,” or “ok,” which illustrated that a process involving 2 different sockets was not a particular issue. For example, 1 participant discussed their experience in terms of the output of the session:

I got something out of it, it looks like this time if you know what I mean. At the end of this fitting there was something that looked like it was ready for me.
[Participant 10, male, 58 y, amputation following trauma, gel cushion liner]

Other participants discussed how they felt about comparing the 2 sockets and commented on how they did not know who designed either of the 2 sockets, for example:

I didn’t know who designed and who produced each socket.
[Participant 1, male, 81 y, infection, sock]

Other participants discussed their experience in terms of their expectations of addressing their specific socket fit issues relating to their residual limb. For example:

The only thing I’ve got hope for there is that the bottom of my stump shrinks a lot more before it gets thinner and thinner because that’s the only thing that’s stopping me having a smaller prosthetic is that stump. It’s where they added all the bits on.
[Participant 10, male, 58 y, amputation following trauma, gel cushion liner]
Theme 2: Similarity Between the Compared Sockets

Several participants described not feeling any difference in the level of comfort between the sockets, regardless of which design process was used. For some participants, this appeared to be related to feeling that both sockets were equally comfortable, with descriptions of how they found it difficult to decide if one was more comfortable than the other:

As far as comfort there was very little difference in it really to be honest as far as I could judge at the time.
[Participant 19, male, 71 y, amputation due to ischemia, gel liner]

Some participants compared the comfort of the two new sockets to their previous device. For example:

I feel a lot more comfortable really than the old one.
[Participant 1, male, 81 y, infection, sock]

This similarity in the level of comfort between the 2 new sockets was described by some participants in relation to both sockets feeling equally “firm,” “stable,” and “both fitting well,” and both sockets feeling very “natural,” as described by 1 participant:

Absolutely fine. They were both very, very close to being completely natural.
[Participant 4, male, 30 y, elective amputation following trauma, suction liner]

However, for other participants, the similarity between the sockets appeared to be associated with feeling similar levels of pain:

...walking with them as I was walking there was the same amount of pain...
[Participant 11, female, 84 y, peripheral vascular disease, cushion liner]
Theme 3: Differences Between the Compared Sockets

Other participants noticed significant differences in the comfort of the 2 sockets, from positive and negative perspectives. The reported comparative fitment assessments were associated with some of the factors the participants raised in theme 2, around similarities in fit, but also included differing, design-related factors such as the height of the socket, local shape details associated with bony prominences, and dynamic fit. Participants mentioned issues such as pressure points, discomfort at the back of the leg, and feelings of tightness. For example:

One felt really tight. I like it tight when I get a new socket because it lasts longer. One felt a lot more comfortable at the front where my bone is closer to the skin.
[Participant 12, male, 40 y, amputation due to chronic ulcers, seal-in liner]

Differences in the size of the sockets also affected comfort. One participant explained:

The taller one was a bit more uncomfortable due to how high it was. When you bend your leg down, it hits the back of your leg.
[Participant 4, male, 30 y, elective amputation post trauma, suction liner]

Some participants felt that the height of the socket made it feel more natural. These differences impacted participants’ walking, with 1 socket feeling easier to walk in than the other

One socket was absolutely perfect. The other one had different movement, slightly different, but it seemed like you walked quicker with it. It was a nice, easy movement.
[Participant 8, male, 76 y, ischemic amputation, pin liner]
Theme 4: Residual Limb Factors Impacting Perceptions of Socket Comfort

While comparing the 2 sockets, participants discussed sensitive areas that affected their comfort, such as spots on the crest of the shin and other specific sore areas. These issues were often ongoing and considered by the prosthetist during socket design:

I’ve always had this issue. We’ve had to shave out a bit of the socket to ease the pressure. But I think it’s just the shape of my stump, and it’s something that will always be an issue and I’m always aware of it.
[Participant 13, male, 32 y, elective amputation post trauma, no interface details recorded]

Some participants reflected on the adjustment after the fitting process, saying:

I don’t know if the second one on the right whether it needed any more adjustment. Maybe a little bit but I would say they are fairly comfortable.
[Participant 16/17, male, 70 y, diabetic amputations, no interface details recorded]

Other factors, also not necessarily directly related to the design process, appeared to impact the participants’ views of which of the 2 new sockets were more comfortable. For example, 1 participant described how their residual limb had shrunk since their last prosthetic fitting, and this led to the new socket feeling more comfortable:

But it does feel better and that clear one that I had in my old socket … had more movement because the stump had shrunk so much. So, it’s just a case of getting used to having this because it’s slightly smaller and it’s hugging the stump. Perhaps where I had the freedom before where it moves about a lot more … I’ve got to get used to the new feelings of it.
[Participant 6, male, 51 y, dysvascular amputation, no interface devices]

Principal Results

This study demonstrated the noninferiority of the EG socket design method in comparison with conventional clinician-led CAD/CAM socket design at initial fitting. SCS indicated similar comfort on average and reduced variability across the cohort for the EG sockets. Participant interviews confirmed this assessment and added a more detailed understanding of socket satisfaction.

Thematic analysis also revealed the patients’ detailed understanding of the nuances of their prosthesis design and its fit and demonstrated the importance of ensuring a trained, experienced clinician directs the application of technology-enhanced socket design processes. The comfort results indicated noninferiority of the EG sockets’ fundamental design without any personalized clinician input. However, the qualitative study provided evidence to support clinical usage of EG sockets with an expert human in the loop to make design decisions in response to patients’ individual and complex needs [12,24,37]. That might include local design modifications in response to vulnerable sites on that individual’s residual limb or accommodating the patient’s preferences for tightness of fit. This was identified for the single participant whose EG socket was rated with SCS below 7 (participant 16/17, male, 70 y, diabetic amputation), for whom the score of 7 was attributed to an unusual tissue sensitivity at a supracondylar site, where the CAD/CAM control socket featured a local modification. The same point may explain the observed trend where all 4 EG sockets that were rated as less comfortable than the CAD/CAM control were cases where the control scored an SCS of 9 to 10, where such personalization has evidently been successful. Patient-specific local areas of sensitivity and corresponding design changes inherently cannot be generalized, so this justifies the importance of an expert in the design process, who retains clinical responsibility.

Comparison With Previous Work

The complementary evidence provided by this multiple methods approach demonstrates the value of exploring the participants’ experience more deeply through interviewing, to enhance what can be learned from objective measurement [38]. However, this approach is somewhat unusual in the assessment of prosthetic technologies [30]. The study involved a large interdisciplinary team, which enabled separation of trial socket design, clinical assessment, interviewing, data analysis, and interpretation, in an attempt to minimize potential researcher biases. The combined synthesis of the SCS and interviews revealed the participants’ detailed understanding of the influence of residual limb shape and corresponding socket design upon their comfort and function, illustrative of the value of excellent prosthetist-patient communication. These technologies may offer an opportunity to improve patient experience through understanding of their care [39] or even shared decision-making [40]. The use of an EG socket design process may enable greater focus on the higher value-added aspects of personalization and might be easier to perform in front of the patient [24]. Most notably, this observation justifies ongoing work with clinician stakeholders, for example, to indicate specific areas in which human input is particularly important for fitting EG sockets to newly presenting individuals, and in the development of software interfaces, building upon methods reported by Ngan et al [41].

Limitations

The study included a relatively diverse group of patients accessing prosthetic rehabilitation services, except for gender, where the recruited participants were predominantly men (16/17) [42], which may impact the study’s generalizability. Women are less likely to have a major amputation and to be successfully fitted with a prosthesis [43], enter prosthetic rehabilitation later [44], and are underrepresented in research cohorts [45]. It is not clear why this study’s convenience sampling resulted in an imbalance, but this illustrates the need to ensure that further studies include more diverse gender representation, as the experiences of men and women may be different. However, the distribution of randomly sampled people whose socket designs were used to train the socket design evidence model [27] was representative of the population of people with transtibial amputations, and external validity is supported by their diverse range of ages, reasons for limb absence, activity levels, and time since amputation. Further bias may arise from the recruited participants’ age profile (median 66, IQR 51‐73, range 30‐84 y), which was similar on average but narrower in range than the general population of people using prosthetic limbs [42]. The participants' age profile was older on average but similar in range to the historic socket design dataset (median 60, IQR 50‐71, range 20‐94 y [27]). Finally, the training datasets came from clinicians at a single center, whose practice may be similar, potentially limiting the diversity of design approaches.

The study also has limitations in its single timepoint and outcome measure, where previous studies comparing CAD/CAM with plaster design and fabrication considered quality of life [7] and more extensive assessments of comfort, fit, cosmesis, weight, and function alongside clinical workload and productivity measures [9]. The study cannot indicate how socket comfort would develop long-term, and it may be preferable to use the expanded SCS for best, worst, and average comfort over a longer period [46]. However, this study replicates the clinical assessment of trial and definitive sockets in UK National Health Service settings and provides additional insight since participants could compare 2 socket options side-by-side and select their preference, with interviews proceeding a month after fitting. A different comparison, versus plaster socket design methods, might provide additional insights, but the study was designed to compare with conventional CAD/CAM as the standard of care in the participating centers, and indeed, the prevalence of CAD/CAM is rising. Finally, the study was not formally randomized, and prosthetists generally fitted their own design socket first, meaning a carryover effect cannot be excluded. Alongside this limitation, single-blinding could not apply to 3 participants for whom the EG check socket was compared with a definitive CAD/CAM control socket, instead of comparing 2 visually-similar check sockets.

Clinical Messages

First, EG transtibial prosthetic socket designs compared well with devices produced by conventional CAD/CAM clinical practice, in terms of comfort at socket fitting and patient feedback in semistructured interviews.

Second, qualitative feedback confirmed that clinician input remains essential to incorporate patient-specific socket design details in response to local sites of sensitivity or tissue vulnerability, or preference: details that cannot be generalized.

Finally, work is underway to incorporate EG socket designs into CASD software such that they can be modified at fitting, enhancing evidence-based practice as a support tool to help qualified clinicians leverage their experience and skill and enable co-design with the prosthesis user.

Conclusions

Overall, the study findings support wider clinical use of EG transtibial sockets, but also demonstrate the importance of delivering this technology in a way that facilitates prosthetist input in tailoring the design to their individual patients’ needs and provides maximal opportunity for their communication. Such an implementation may enable prosthetists to take the advantages offered by technology while retaining or ideally enhancing trust through human-centered prosthetic rehabilitation.

Acknowledgments

We acknowledge our study participants, patient advisor the Alex Lewis Trust, and kind support from the University of Southampton’s Institute for Life Sciences. This research was funded by the Royal Academy of Engineering (RAEng; grants RF/130 and EF1819\8\24), Innovate UK (grant 10014827), the Higher Education Innovation Fund (HEIF), and the Alan Turing Institute (grant EP/N510129/1). The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, or in the decision to publish the results.

Data Availability

The dataset supporting the conclusions of this paper is available in the University of Southampton repository [47].

Authors' Contributions

FM contributed to data collection, data analysis, data interpretation, writing, and editing. MD-H contributed to study design, literature search, data analysis, data interpretation, writing, and editing. JB contributed to study design, data collection, data analysis, data interpretation, and editing. JS contributed to study design, technical design, data analysis, data interpretation, figures, writing, and editing. CR contributed to technical design, data analysis, underlying data verification, and editing. PW, CO, CM, and DH contributed to study design, data interpretation, and editing. CW, JK, SS, KM, SG, HH, DHS, and VK contributed to study design, data collection, data interpretation, and editing. AD contributed to study design, technical design, literature search, data interpretation, figures, writing, and editing.

Conflicts of Interest

Authors FM, MD-H, CO, CM, SG, HH, and VK have no conflict of interest to declare. Authors JS, JB, CR, AD, and PW declare employment and shareholding in Radii Devices Ltd. Authors DH, CW, JK, SS, KM, and DHS declare employment at Opcare Ltd.

Multimedia Appendix 1

Participant consent form.

DOCX File, 456 KB

Multimedia Appendix 2

Transtibial checkout procedure.

DOCX File, 25 KB

Multimedia Appendix 3

Semistructured interview schedules.

DOCX File, 28 KB

  1. Saunders CG, Foort J, Bannon M, Lean D, Panych L. Computer aided design of prosthetic sockets for below-knee amputees. Prosthetics & Orthotics International. 1985;9(1):17-22. [CrossRef]
  2. Köhler P, Lindh L, Netz P. Comparison of CAD-CAM and hand made sockets for PTB prostheses. Prosthet Orthot Int. Apr 1989;13(1):19-24. [CrossRef] [Medline]
  3. Topper AK, Fernie GR. An evaluation of computer aided design of below-knee prosthetic sockets. Prosthet Orthot Int. Dec 1990;14(3):136-142. [CrossRef] [Medline]
  4. Ruder GK. CAD CAM trans-tibial temporary prosthesis: analysis and comparison with an established technique. Prosthet Orthot Int. Dec 1992;16(3):189-195. [CrossRef] [Medline]
  5. Oberg T, Lilja M, Johansson T, Karsznia A. Clinical evaluation of trans-tibial prosthesis sockets: a comparison between CAD CAM and conventionally produced sockets. Prosthet Orthot Int. Dec 1993;17(3):164-171. [CrossRef] [Medline]
  6. Wong MS, Cheng JCY, Wong MW, So SF. A work study of the CAD/CAM method and conventional manual method in the fabrication of spinal orthoses for patients with adolescent idiopathic scoliosis. Prosthet Orthot Int. Apr 2005;29(1):93-104. [CrossRef] [Medline]
  7. Karakoç M, Batmaz İ, Sariyildiz MA, Yazmalar L, Aydin A, Em S. Sockets manufactured by CAD/CAM method have positive effects on the quality of life of patients with transtibial amputation. Am J Phys Med Rehabil. Aug 2017;96(8):578-581. [CrossRef] [Medline]
  8. van der Stelt M, Berends B, Papenburg M, et al. Evaluating the effectiveness of transtibial prosthetic socket shape design using artificial intelligence: a clinical comparison with traditional plaster cast socket designs. Arch Phys Med Rehabil. Feb 2025;106(2):239-246. [CrossRef] [Medline]
  9. Houston VL, Burgess EM, Childress DS, et al. Automated fabrication of mobility aids (AFMA): below-knee CASD/CAM testing and evaluation program results. J Rehabil Res Dev. 1992;29(4):78-124. [CrossRef] [Medline]
  10. Smith DG, Burgess EM. The use of CAD/CAM technology in prosthetics and orthotics--current clinical models and a view to the future. J Rehabil Res Dev. 2001;38(3):327-334. [Medline]
  11. Dean D, Saunders CG. A software package for design and manufacture of prosthetic sockets for transtibial amputees. IEEE Trans Biomed Eng. Apr 1985;32(4):257-262. [CrossRef] [Medline]
  12. Boone DA, Harlan JS, Burgess EM. Automated fabrication of mobility aids: review of the AFMA process and VA/Seattle ShapeMaker software design. J Rehabil Res Dev. 1994;31(1):42-49. [Medline]
  13. Lemaire ED, Johnson F. A quantitative method for comparing and evaluating manual prosthetic socket modifications. IEEE Trans Rehabil Eng. Dec 1996;4(4):303-309. [CrossRef] [Medline]
  14. Cutti AG, Santi MG, Hansen AH, Fatone S, Residual Limb Shape Capture Group. Prosthetist-specific rectification templates based on artificial intelligence for the digital fabrication of custom transtibial sockets. Prosthesis. 2024;6(5):1149-1169. [CrossRef]
  15. Torres-Moreno R, Foort J, Morrison JB, Saunders CG. A reference shape library for computer aided socket design in above-knee prostheses. Prosthet Orthot Int. Dec 1989;13(3):130-139. [CrossRef] [Medline]
  16. Fatone S, Johnson WB, Tran L, Tucker K, Mowrer C, Caldwell R. Quantification of rectifications for the Northwestern University Flexible Sub-Ischial Vacuum Socket. Prosthet Orthot Int. Jun 2017;41(3):251-257. [CrossRef] [Medline]
  17. Weiss HR, Werkmann M. "Brace Technology" thematic series - the ScoliOlogiC® Chêneau light™ brace in the treatment of scoliosis. Scoliosis. Sep 6, 2010;5:1-18. [CrossRef] [Medline]
  18. Bowman GG. The L.A. Brace-description and discussion. Physiother Theory Pract. Jan 2011;27(1):68-73. [CrossRef] [Medline]
  19. Li S, Lan H, Luo X, Lv Y, Gao L, Yu H. Quantitative compensation design for prosthetic socket based on eigenvector algorithm method. Rev Sci Instrum. Oct 1, 2019;90(10). [CrossRef]
  20. Karamousadakis M, Porichis A, Ottikkutti S, Chen D, Vartholomeos P. A sensor-based decision support system for transfemoral socket rectification. Sensors (Basel). May 28, 2021;21(11):3743. [CrossRef] [Medline]
  21. Gu Y, He L, Zeng H, et al. A data-driven design framework for structural optimization to enhance wearing adaptability of prosthetic hands. IEEE Trans Neural Syst Rehabil Eng. 2024;32:2621-2632. [CrossRef] [Medline]
  22. Lee DRC, Yang X, Riccio-Ackerman F, et al. A clinical comparison of a digital versus conventional design methodology for transtibial prosthetic interfaces. Sci Rep. Oct 28, 2024;14. [CrossRef] [Medline]
  23. Oldfrey BM, Morgado Ramirez DZ, Miodownik M, et al. A scoping review of digital fabrication techniques applied to prosthetics and orthotics: Part 1 of 2—Prosthetics. Prosthetics & Orthotics International. 2024;48(5):574-589. [CrossRef]
  24. Dickinson A, Diment L, Morris R, Pearson E, Hannett D, Steer J. Characterising residual limb morphology and prosthetic socket design based on expert clinician practice. Prosthesis. 2021;3(4):280-299. [CrossRef]
  25. Nagarajan YR, Farukh F, Silberschmidt VV, Kandan K, Singh AK, Mukul P. Shape analysis of prosthetic socket rectification procedure for transtibial amputees. Prosthesis. 2024;6(1):157-174. [CrossRef]
  26. Ngan CC, Pendse V, Sivasambu H, Ouellette E, Ready N, Andrysek J. Preliminary characterization of rectification for transradial prosthetic sockets. Sci Rep. Mar 8, 2024;14(1):5759. [CrossRef] [Medline]
  27. Dickinson AS, Steer JW, Rossides C, et al. Insights into the spectrum of transtibial prosthetic socket design from expert clinicians and their digital records. Front Rehabil Sci. 2024;5:1354069. [CrossRef] [Medline]
  28. von Elm E, Altman DG, Egger M, et al. The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Bull World Health Organ. Nov 2007;85(11):867-872. [CrossRef] [Medline]
  29. The Voices project. People Powered Prosthetics. 2024. URL: https://www.peoplepoweredprosthetics.com/projects/voices/ [Accessed 2025-07-18]
  30. Sanders JE, Vamos AC, Mertens JC, et al. An adaptive prosthetic socket for people with transtibial amputation. Sci Rep. May 15, 2024;14(1):11168. [CrossRef] [Medline]
  31. Turner S, McGregor AH. Perceived effect of socket fit on major lower limb prosthetic rehabilitation: a clinician and amputee perspective. Arch Rehabil Res Clin Transl. Sep 2020;2(3):100059. [CrossRef] [Medline]
  32. Anderson CB, Kittelson AJ, Wurdeman SR, et al. Understanding decision-making in prosthetic rehabilitation by prosthetists and people with lower limb amputation: a qualitative study. Disabil Rehabil. Feb 2023;45(4):723-732. [CrossRef] [Medline]
  33. Gavette H, McDonald CL, Kostick-Quenet K, Mullen A, Najafi B, Finco MG. Advances in prosthetic technology: a perspective on ethical considerations for development and clinical translation. Front Rehabil Sci. 2023;4:1335966. [CrossRef] [Medline]
  34. Hafner BJ, Morgan SJ, Askew RL, Salem R. Psychometric evaluation of self-report outcome measures for prosthetic applications. J Rehabil Res Dev. 2016;53(6):797-812. [CrossRef] [Medline]
  35. Efron B. Censored data and the bootstrap. J Am Stat Assoc. 1981;76(374):312-319. [CrossRef]
  36. Braun V, Clarke V. Using thematic analysis in psychology. Qual Res Psychol. Jan 2006;3(2):77-101. [CrossRef]
  37. Foort J. Innovation in prosthetics and orthotics. Prosthet Orthot Int. 1986;10(2):61-71. [CrossRef]
  38. Dillon MP, Fatone S, Ramstrand N, Hafner BJ. Prosthetics and Orthotics International welcomes qualitative research submissions. Prosthet Orthot Int. 2019;43(4):366-368. [CrossRef]
  39. Adams RJ. Improving health outcomes with better patient understanding and education. Risk Manag Healthc Policy. 2010;3:61-72. [CrossRef] [Medline]
  40. Anderson CB, Fatone S, Mañago MM, et al. Improving shared decision-making for prosthetic care: a qualitative needs assessment of prosthetists and new lower-limb prosthesis users. Prosthet Orthot Int. Feb 1, 2023;47(1):26-42. [CrossRef] [Medline]
  41. Ngan CC, Sivasambu H, Kelland K, Ramdial S, Andrysek J. Understanding the adoption of digital workflows in orthotic & prosthetic practice from practitioner perspectives: a qualitative descriptive study. Prosthetics & Orthotics International. 2022;46(3):282-289. [CrossRef]
  42. Twiste M. Limbless statistics annual report: 2011/2012. University of Salford Manchester; 2015. URL: https:/​/shop.​salford.ac.uk/​product-catalogue/​university-goods-and-services/​limbless-statistics/​limbless-statistics-annual-report-201112
  43. Davie-Smith F, Paul L, Nicholls N, Stuart WP, Kennon B. The impact of gender, level of amputation and diabetes on prosthetic fit rates following major lower extremity amputation. Prosthet Orthot Int. Feb 2017;41(1):19-25. [CrossRef] [Medline]
  44. Levy J, Kamdar N, Adidharma W, Kemp S, Hooper R. 250 gender disparities in the acquisition of lower extremity prosthetics following major limb amputation. J Clin Trans Sci. Apr 2024;8(s1):75-75. [CrossRef]
  45. Ljung J, Iacangelo A. Identifying and acknowledging a sex gap in lower-limb prosthetics research: a review. J Prosthet Orthot. 2024;36(1):e18-e24. [CrossRef]
  46. Morgan SJ, Askew RL, Hafner BJ. Measurements of best, worst, and average socket comfort are more reliable than current socket comfort in established lower limb prosthesis users. Arch Phys Med Rehabil. Jun 2022;103(6):1201-1204. [CrossRef] [Medline]
  47. Mbithi F, Donovan-Hall M, Bramley JL, et al. Evidence-generated sockets for transtibial prosthetic limbs: a multiple-methods study from the patient’s perspective, compared to conventional computer-aided designs. University of Southampton Institutional Repository. URL: https://eprints.soton.ac.uk/493705/ [Accessed 2025-07-25]


CAD/CAM: computer-aided design and manufacturing
CASD: computer-aided socket design
EG: evidence-generated
PPIE: patient and participant involvement and engagement
SCS: socket comfort score
STROBE: Strengthening the Reporting of Observational Studies in Epidemiology


Edited by Sarah Munce; submitted 16.12.24; peer-reviewed by Ankit Sahai, Calvin Ngan, L M Verhamme; final revised version received 04.04.25; accepted 01.05.25; published 21.08.25.

Copyright

© Florence Mbithi, Maggie Donovan-Hall, Jennifer Bramley, Joshua Steer, Charalambos Rossides, Peter Worsley, Chantel Ostler, Cheryl Metcalf, Dominic Hannett, Caroline Ward, Jack Kitchen, Sioned Steventon, Katy McIntosh, Shigong Guo, Helen Harvey, David Henderson Slater, Vijay Kolli, Alex Dickinson. Originally published in JMIR Rehabilitation and Assistive Technology (https://rehab.jmir.org), 21.8.2025.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Rehabilitation and Assistive Technology, is properly cited. The complete bibliographic information, a link to the original publication on https://rehab.jmir.org/, as well as this copyright and license information must be included.