Original Paper
Abstract
Background: In recent years, there has been a great interest in analyzing upper-limb kinematics. Inertial measurement with mobile phones is a convenient and portable analysis method for studying humerus kinematics in terms of angular mobility and linear acceleration.
Objective: The aim of this analysis was to study upper-limb kinematics via mobile phones through six physical properties that correspond to angular mobility and acceleration in the three axes of space.
Methods: This cross-sectional study recruited healthy young adult subjects. Humerus kinematics was studied in 10 young adults with the iPhone4. They performed flexion and abduction analytical tasks. Mobility angle and lineal acceleration in each of its axes (yaw, pitch, and roll) were obtained with the iPhone4. This device was placed on the right half of the body of each subject, in the middle third of the humerus, slightly posterior. Descriptive statistics were calculated.
Results: Descriptive graphics of analytical tasks performed were obtained. The biggest range of motion was found in pitch angle, and the biggest acceleration was found in the y-axis in both analytical tasks. Focusing on tridimensional kinematics, bigger range of motion and acceleration was found in abduction (209.69 degrees and 23.31 degrees per second respectively). Also, very strong correlation was found between angular mobility and linear acceleration in abduction (r=.845) and flexion (r=.860).
Conclusions: The use of an iPhone for humerus tridimensional kinematics is feasible. This supports use of the mobile phone as a device to analyze upper-limb kinematics and to facilitate the evaluation of the patient.
doi:10.2196/rehab.4101
Keywords
Introduction
Upper-limb mobility is of great interest in clinical settings [
] because measuring the range of motion (ROM) is critical when evaluating the musculoskeletal system [ ]. Upper extremities have been measured by manual goniometry for the last 100 years, but measurement methods have recently expanded [ - ]. Besides goniometry, arm ROM has been studied by other methods, such a digital goniometer [ ], visual estimation [ ], digital inclinometer [ , ], three-dimensional (3D) gyroscope [ , ], polhemus fastrak [ ], calibration anatomical system techniques [ ], the Kinect system [ ], biplane fluoroscopy [ , ], markers fitted on intracortical pins [ ], 3D computerized tomography [ ], and the moiré fringe projection technique [ ].Recently, telerehabilitation has provided rehabilitation using Internet communication as a result of emerging contemporary technologies for therapeutic purposes [
, ]. Thus, Internet-based evaluation and goniometry have been accepted as new, valid, and reliable tools for measuring ROM [ ]. This drives the use of the mobile phone as a tool for assessing and measuring. Mobile phone apps are being validated as goniometric tools [ ] through clinometers [ ] or goniometers [ ]. Image-based apps have been created for measuring elbow and hallux valgus angles [ , ], and clinometer-based apps have also been created for measuring shoulder ROM [ ]. In addition, an inclinometer-based app on a mobile phone has been demonstrated to have an acceptable reliability compared to conventional inclinometers that evaluate the shoulder joint [ ]. Furthermore, active shoulder external rotation measures have been validated using inclinometery-based and image-based apps [ ]. Recently, a study has analyzed arm motion by inertial variables provided by a mobile phone in five subjects [ ].One of the recently used techniques has been inertial sensors. Their use in human analysis involves a valid and reliable method that provides the potential required for dynamic 3D motion analysis [
]. Their protocol [ ] and intra- and interoperator reliability [ ] in the upper extremity have been determined. In addition, their operational feasibility in various clinical applications has been studied [ ]. Several protocols have also been developed for analyzing the scapulothoracic, humerothoracic, and elbow joints [ ], as well as scapula [ ]. Very recently, reliability and precision of scapula kinematic through inertial and magnetic measurement systems (IMMS) has been studied in healthy subjects [ ]. Advantages and disadvantages of these sensors have been discussed as part of a variety of motion analysis systems [ ]. Thus, inertial sensors embedded in mobile phones have been used for analyzing movement, such as trunk kinematics [ ]. More specifically, they have been used for evaluating shoulder movement using kinematic scores to assess the difference between healthy and painful shoulders [ ].Emerging mobile phone use for therapeutic purposes [
] has led to the need for research on arm ROM using mobile phones while incorporating the qualities of inertial sensors that allow clinicians an inexpensive and easy-to-use tool for upper extremity evaluation and outcome assessment.The purpose of this study was to study humerus kinematics through two physical properties that correspond to angular mobility and acceleration in the three axes of space, obtained by inertial sensors embedded in a mobile phone.
Methods
Subjects
This cross-sectional study recruited healthy young adults from the Faculty of Health Sciences (University of Málaga) who were interested in taking part in the project. Subjects provided inclusion and exclusion criteria. Inclusion criteria included being aged between 18 and 35 years, having a Body Mass Index (BMI) between 18.5 and 28, and being right-handed. Exclusion criteria included consuming analgesics or non-steroidal anti-inflammatory drugs (NSAIDs) and suffering from shoulder pathology.
Ten subjects (7 men and 3 women) were included. Mean age was 24.2 years (SD 4.04 years), and average BMI was 22.59 kg/m2(SD 2.4 kg/m2; see
).The ethics committee of the University of Málaga, Spain, approved this study. Written consent was obtained following an explanation of the procedures.
Minimum | Maximum | Mean | Standard deviation | |
Age, years | 20.00 | 34.00 | 24.20 | 4.04 |
Size, cm | 156.00 | 184.00 | 172.20 | 9.05 |
Weight, kg | 48.00 | 87.00 | 66.60 | 11.88 |
BMI, kg/m2 | 19.72 | 27.46 | 22.59 | 2.40 |
Apparatus
Mobility angle (degrees) and acceleration were measured along three orthogonal axes using the iPhone4 (LG Electronics INC, Seoul, South Korea) iOS8.2, which has a storage capacity of 20MB. This phone was placed on the right half of the body of each subject in the triceps skinfold site, located on the posterior part of triceps at mid-acromiale-radiale level (defined by ISAK) [
]. The phone was attached by using a neoprene arm belt ( ) and remained attached throughout. The app used to obtain kinematic data was xSensor Pro (Crossbow Technology, Inc.), available at the Apple AppStore. The data-sampling rate was set to 32 Hz, and the data for each analytical task was transmitted as email for analysis and post-processing. Data from the phone were subsequently sent to a Microsoft Excel 2007 database.Because of its positioning, axes and planes in the phone corresponded to different planes of anatomical movement: yaw (z) for shoulder flexor-extension plane, pitch (y) for shoulder abduction plane, and roll (x) for humerus rotation plane (
).Procedure
Subjects were asked to attend the study in the Human Movement Laboratory, Faculty of Health Sciences (University of Málaga). The analytical task to be performed was explained clearly. The beginning and the end were decided by a verbal order by the researcher, which was identical for all participants. They stood, starting from a neutral position, and performed the following analytical tasks: right shoulder abduction for eight repetitions and, after a break of about three minutes, right shoulder flexion for eight repetitions. Participants were told to perform the movements to the highest position they could reach. Both tasks were performed with the elbow extended, the wrist in a neutral position, and the palm area of the hand toward the midline at the beginning and end of the movement.
Data Analysis
SPSS v15.0 was used for all statistical computations. Descriptive statistics (mean, standard deviation, minimum, and maximum) were calculated for age, height, weight, BMI, angular mobility, and linear acceleration. Standard procedures were used to calculate means and standard deviations. The Kolmogorov-Smirnov test showed a normal distribution of the data (P>.05).
Angular mobility and linear acceleration were calculated in two different ways: calculating each space of motion separately and considering the resultant vector of the three axes of movement, which was understood as: Resultant vector = √x2+ y2+z2
Results
Analyzing angular mobility allowed us to obtain descriptive graphics of analytical tasks performed by each participant (
).Means and standard deviations of angular mobility and acceleration were calculated. For that, data from the second repetition of the second series for both abduction and flexion movements in each of the space axes were analyzed.
In terms of angular mobility, the biggest range was found in pitch axis for flexion movement, followed by the same axis in abduction. However, the smallest range was found in yaw for flexion and in roll for abduction. Considering resultant vector, ROM is bigger for abduction (
).Regarding acceleration, the largest value was found in the y-axis, followed by the z- and x-axes in both movements. Flexion acceleration was greater in the x-axis when compared to abduction; while, in abduction, acceleration was greater in the y- and z-axes than flex. With regards to resultant vector, acceleration was greater in abduction (
).Relationship between angular mobility and linear acceleration was calculated for both tasks in each axes of space and resultant vector. Strong correlation was found in y and x as well as in resultant vector, for both tasks. However, that correlation was not significant in yaw axis. More details are shown in
.Angles | Abduction, mean (SD) | Flexion, mean (SD) |
Yaw | 109.96 (39.44) | 79.81 (39.69) |
Pitch | 151.59 (10.21) | 156.15 (12.40) |
Roll | 87.53 (38.46) | 80.09 (47.45) |
Resultant vector | 209.69 (42.01) | 197.89 (42.02) |
Axes | Abduction, mean (SD) | Flexion, mean (SD) |
X | 8.48 (1.76) | 8.53 (2.8) |
Y | 19.48 (0.85) | 19.43 (0.77) |
Z | 9.41 (1.5) | 7.09 (1.9) |
Resultant vector | 23.31 (1.58) | 22.55 (1.73) |
Task | Abduction, correlation (P value) | Flexion, correlation (P value) |
Yaw (z) | .462 (.17) | .380 (.2) |
Pitch (y) | .914 (<.01) | .915 (<.01) |
Roll (x) | .811 (<.01) | .691 (.02) |
Resultant vector | .845 (<.01) | .860 (<.01) |
Discussion
Principal Findings
This study has described and examined upper-limb 3D kinematics using the inertial sensor in the iPhone4 during the performance of shoulder abduction and flexion task in healthy subjects. The biggest range of motion and largest acceleration values varied along each axes. However, taking into account resultant vector from each axes of space, mobility and acceleration were found to be greater for abduction movement. Strong correlation was found between tridimensional mobility and acceleration for both task. The results obtained in this study allow us to obtain descriptive data from upper-limb 3D kinematics, providing an overview of the use of a mobile phone for the study of upper-limb movement.
Previous research has attempted to describe overall upper-limb kinematics through mobile phones. Recently, mobile phone inclinometric measurements of various movements, including abduction and flexion, were performed in 41 affected shoulders. Results showed an acceptable reliability score when compared to conventional goniometers [
]. Very recently, functional assessments of the shoulder through velocity and acceleration inertial variables provided by a mobile phone were studied in five subjects [ ]. The use of a clinometer embedded in a mobile phone has been validated in shoulder abduction and flexion movements in healthy and symptomatic shoulders [ ]. Recently, mobile phone goniometric measurements have been validated in five healthy subjects, obtaining 95.2° for flexion and 155.4° for abduction [ ], which is similar to the abduction degrees obtained in our study. Furthermore, an inclinometery-based and photo-based mobile phone app has been validated for measuring shoulder external rotation [ ].Upper-limb motion has been studied using several devices from decades ago. However, it tends to be deep in kinematic aspects [
] and 3D kinematics [ , ]. For that reason, inertial devices have played an important role when studying shoulder kinematics in several studies [ , ]. Obtaining different results depending on analyzing one plane/axis or its resultant vector intensifies the importance of taking into account the three-dimensional component of anatomical movement, whose analysis is allowed through inertial sensors embedded in mobile phones.Nowadays, because of new technologies, the concept of telerehabilitation has emerged as an attractive opportunity for provisioning rehabilitation at a distance with the Internet, thus improving the quality of rehabilitation health care [
, ]. Providing comprehensive instructions regarding placement and use of mobile phones would allow patients to measure humerus kinematics, facilitating equitable access to all individuals. Regarding upper-limbs, diagnosis and assessment of musculoskeletal shoulder disorders through the Internet have already been studied [ ]. As telerehabilitation is a convenient and easy-to-use system, it would help patients and physicians meet health-related goals. Communication technologies as part of telehealth should also reduce health care costs.Having reference values of humerus kinematics in the future would be potentially desirable for comparing data from new technologies, such smartphones or smartcameras like Kinect, opening a new world of possibilities in shoulder telehealth assessment.
Tridimensional kinematic tendency, along with the birth of the concept of telerehabilitation, shows the need for mobile phone 3D evaluation of arm movements. The results of this study are in line with other research and show that the use of inertial sensors embedded in mobile phones for upper-limb kinematic analysis appears feasible.
Limitations
The main weakness of the study is that it is a cross-sectional study, which means cause and effect relationships in kinematic patterns cannot be established. In addition, criterion validity has not been studied because there is no criterion standard.
However, having a sample with a larger number of participants and in which there are also subjects presenting shoulder pathology, we hope to compare our results with those studies reporting on other systems for upper-limb motion analysis. Furthermore, measuring upper-limbs with a gold standard system will allow us to validate mobile phones in upper-limb use as an inertial, easy-to-use measurement. It should be also mentioned that this study estimated only humerus kinematics, while the contribution of other shoulder joints, like sternoclavicular and acromioclavicular ones [
], whose importance has been previously claimed were not included.Conclusion
This study discusses humerus kinematics and identifies movement patterns. Therefore, it supports using mobile phones as devices to analyze upper-limb kinematics. Thanks to this study, it is possible to develop a simple and accessible-to-all app that facilitates patient evaluation in this area.
Acknowledgments
This study received a grant for Best Protocol Research in 2013 from the Chartered Society of Physiotherapy in Andalusia, Spain.
The authors are grateful to the volunteers for their participation.
Authors' Contributions
AICV and PB participated in the conception and design of the study; the data collection, analysis, and interpretation of data; and helped to draft the manuscript. CRJ participated in the data collection, analysis, and interpretation of data and drafted the manuscript. All authors read and approved the final manuscript.
Conflicts of Interest
None declared.
References
- Muir SW, Luciak Corea C, Beaupre L. Evaluating change in clinical status: reliability and measures of agreement for the assessment of glenohumeral range of motion. N Am J Sports Phys Ther 2010 Sep;5(3):98-110 [FREE Full text] [Medline]
- McFarland EG. Examination of the Shoulder: The Complete Guide. New York: Thieme; 2006.
- Boone DC, Azen SP, Lin CM, Spence C, Baron C, Lee L. Reliability of goniometric measurements. Phys Ther 1978 Nov;58(11):1355-1360. [Medline]
- Fortescue Fox R. Physical Remedies for Disabled Soldiers. London: Baillière, Tindall and Cox; 1917.
- Gajdosik RL, Bohannon RW. Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. Phys Ther 1987 Dec;67(12):1867-1872 [FREE Full text] [Medline]
- Rosén NG, Neil G. A simplified method of measuring amplitude of motion in joints. J Bone Joint Surg Arm 1922 Jul 1;4(3):570-579.
- Terwee CB, de Winter AF, Scholten RJ, Jans MP, Devillé W, van SD, et al. Interobserver reproducibility of the visual estimation of range of motion of the shoulder. Arch Phys Med Rehabil 2005 Jul;86(7):1356-1361. [Medline]
- Kolber MJ, Hanney WJ. The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: a technical report. Int J Sports Phys Ther 2012 Jun;7(3):306-313 [FREE Full text] [Medline]
- Kolber MJ, Vega F, Widmayer K, Cheng MSS. The reliability and minimal detectable change of shoulder mobility measurements using a digital inclinometer. Physiother Theory Pract 2011 Feb;27(2):176-184. [CrossRef] [Medline]
- El-Zayat BF, Efe T, Heidrich A, Anetsmann R, Timmesfeld N, Fuchs-Winkelmann S, et al. Objective assessment, repeatability, and agreement of shoulder ROM with a 3D gyroscope. BMC Musculoskelet Disord 2013;14:72 [FREE Full text] [CrossRef] [Medline]
- El-Zayat BF, Efe T, Heidrich A, Wolf U, Timmesfeld N, Heyse TJ, et al. Objective assessment of shoulder mobility with a new 3D gyroscope--a validation study. BMC Musculoskelet Disord 2011;12:168 [FREE Full text] [CrossRef] [Medline]
- Namdari S, Yagnik G, Ebaugh DD, Nagda S, Ramsey ML, Williams GR, et al. Defining functional shoulder range of motion for activities of daily living. J Shoulder Elbow Surg 2012 Sep;21(9):1177-1183. [CrossRef] [Medline]
- Ceccon S, Ceseracciu E, Sawacha Z, Gatta G, Cortesi M, Cobelli C, et al. Motion analysis of front crawl swimming applying CAST technique by means of automatic tracking. J Sports Sci 2013;31(3):276-287. [CrossRef] [Medline]
- Kurillo G, Han JJ, Obdržálek S, Yan P, Abresch RT, Nicorici A, et al. Upper extremity reachable workspace evaluation with Kinect. Stud Health Technol Inform 2013;184:247-253. [Medline]
- Giphart JE, van der Meijden OAJ, Millett PJ. The effects of arm elevation on the 3-dimensional acromiohumeral distance: a biplane fluoroscopy study with normative data. J Shoulder Elbow Surg 2012 Nov;21(11):1593-1600. [CrossRef] [Medline]
- Giphart JE, Brunkhorst JP, Horn NH, Shelburne KB, Torry MR, Millett PJ. Effect of plane of arm elevation on glenohumeral kinematics: a normative biplane fluoroscopy study. J Bone Joint Surg Am 2013 Feb 6;95(3):238-245. [CrossRef] [Medline]
- Dal Maso F, Raison M, Lundberg A, Arndt A, Begon M. Coupling between 3D displacements and rotations at the glenohumeral joint during dynamic tasks in healthy participants. Clin Biomech (Bristol, Avon) 2014 Nov;29(9):1048-1055. [CrossRef] [Medline]
- Park JY, Hwang JT, Kim KM, Makkar D, Moon SG, Han KJ. How to assess scapular dyskinesis precisely: 3-dimensional wing computer tomography--a new diagnostic modality. J Shoulder Elbow Surg 2013 Aug;22(8):1084-1091. [CrossRef] [Medline]
- Gomes PF, Sesselmann M, Faria CDCM, Araújo PA, Teixeira-Salmela LF. Measurement of scapular kinematics with the moiré fringe projection technique. J Biomech 2010 Apr 19;43(6):1215-1219. [CrossRef] [Medline]
- Russell TG. Telerehabilitation: a coming of age. Aust J Physiother 2009;55(1):5-6. [Medline]
- Theodoros D, Russell T. Telerehabilitation: current perspectives. Stud Health Technol Inform 2008;131:191-209. [Medline]
- Russell T. Goniometry via the internet. Aust J Physiother 2007;53(2):136. [Medline]
- Milani P, Coccetta CA, Rabini A, Sciarra T, Massazza G, Ferriero G. Mobile smartphone applications for body position measurement in rehabilitation: a review of goniometric tools. PM R 2014 Nov;6(11):1038-1043. [CrossRef] [Medline]
- Werner BC, Holzgrefe RE, Griffin JW, Lyons ML, Cosgrove CT, Hart JM, et al. Validation of an innovative method of shoulder range-of-motion measurement using a smartphone clinometer application. J Shoulder Elbow Surg 2014 Nov;23(11):e275-e282. [CrossRef] [Medline]
- Kim YS, Jang S, Yoo Y. Mobile Assessment System for Shoulder Joint Rehabilitation: System Development and Preliminary Study. IJBSBT 2014 Apr 30;6(2):51-60. [CrossRef]
- Ferriero G, Sartorio F, Foti C, Primavera D, Brigatti E, Vercelli S. Reliability of a new application for smartphones (DrGoniometer) for elbow angle measurement. PM R 2011 Dec;3(12):1153-1154. [CrossRef] [Medline]
- Walter R, Kosy JD, Cove R. Inter- and intra-observer reliability of a smartphone application for measuring hallux valgus angles. Foot Ankle Surg 2013 Mar;19(1):18-21. [CrossRef] [Medline]
- Shin SH, Ro DH, Lee OS, Oh JH, Kim SH. Within-day reliability of shoulder range of motion measurement with a smartphone. Man Ther 2012 Aug;17(4):298-304. [CrossRef] [Medline]
- Mitchell K, Gutierrez SB, Sutton S, Morton S, Morgenthaler A. Reliability and validity of goniometric iPhone applications for the assessment of active shoulder external rotation. Physiother Theory Pract 2014 Oct;30(7):521-525. [CrossRef] [Medline]
- Oïhénart L, Duc C, Aminian K. iShould: Functional evaluation of the shoulder using a Smartphone. Gait & Posture 2012 Jun;36:S61-S62. [CrossRef]
- Cuesta-Vargas AI, Galán-Mercant A, Williams JM. The use of inertial sensors system for human motion analysis. Phys Ther Rev 2010 Dec;15(6):462-473 [FREE Full text] [CrossRef] [Medline]
- Kontaxis A, Cutti AG, Johnson GR, Veeger HE. A framework for the definition of standardized protocols for measuring upper-extremity kinematics. Clin Biomech (Bristol, Avon) 2009 Mar;24(3):246-253. [CrossRef] [Medline]
- Parel I, Cutti AG, Fiumana G, Porcellini G, Verni G, Accardo AP. Ambulatory measurement of the scapulohumeral rhythm: intra- and inter-operator agreement of a protocol based on inertial and magnetic sensors. Gait Posture 2012 Apr;35(4):636-640. [CrossRef] [Medline]
- Wong WY, Wong MS, Lo KH. Clinical applications of sensors for human posture and movement analysis: a review. Prosthet Orthot Int 2007 Mar;31(1):62-75. [CrossRef] [Medline]
- Cutti AG, Giovanardi A, Rocchi L, Davalli A, Sacchetti R. Ambulatory measurement of shoulder and elbow kinematics through inertial and magnetic sensors. Med Biol Eng Comput 2008 Feb;46(2):169-178. [CrossRef] [Medline]
- Parel I, Cutti AG, Kraszewski A, Verni G, Hillstrom H, Kontaxis A. Intra-protocol repeatability and inter-protocol agreement for the analysis of scapulo-humeral coordination. Med Biol Eng Comput 2014 Mar;52(3):271-282. [CrossRef] [Medline]
- van den Noort JC, Wiertsema SH, Hekman KM, Schönhuth CP, Dekker J, Harlaar J. Reliability and precision of 3D wireless measurement of scapular kinematics. Med Biol Eng Comput 2014 Nov;52(11):921-931. [CrossRef] [Medline]
- Wong WY, Wong MS. Trunk posture monitoring with inertial sensors. Eur Spine J 2008 May;17(5):743-753 [FREE Full text] [CrossRef] [Medline]
- Galán-Mercant A, Cuesta-Vargas AI. Differences in trunk kinematic between frail and nonfrail elderly persons during turn transition based on a smartphone inertial sensor. Biomed Res Int 2013;2013 [FREE Full text] [CrossRef] [Medline]
- Russell TG, Jones AF. Implications of regulatory requirements for smartphones, gaming consoles and other devices. J Physiother 2011;57(1):5-7 [FREE Full text] [CrossRef] [Medline]
- Stewart A, Marfell-Jones M. International Standards for Anthropometric Assessment. In: International Society for the Advancement of Kinanthropometry. New Zealand: Lower Hutt; 2011.
- Cutti AG, Chadwick EK. Shoulder biomechanics and the success of translational research. Med Biol Eng Comput 2014 Mar;52(3):205-210. [CrossRef] [Medline]
- Jolles BM, Duc C, Coley B, Aminian K, Pichonnaz C, Bassin JP, et al. Objective evaluation of shoulder function using body-fixed sensors: a new way to detect early treatment failures? J Shoulder Elbow Surg 2011 Oct;20(7):1074-1081. [CrossRef] [Medline]
- Körver RJP, Heyligers IC, Samijo SK, Grimm B. Inertia based functional scoring of the shoulder in clinical practice. Physiol Meas 2014 Feb;35(2):167-176. [CrossRef] [Medline]
- Steele L, Lade H, McKenzie S, Russell TG. Assessment and Diagnosis of Musculoskeletal Shoulder Disorders over the Internet. Int J Telemed Appl 2012;2012:945745 [FREE Full text] [CrossRef] [Medline]
- Robert-Lachaine X, Marion P, Godbout V, Bleau J, Begon M. Elucidating the scapulo-humeral rhythm calculation: 3D joint contribution method. Comput Methods Biomech Biomed Engin 2015;18(3):249-258. [CrossRef] [Medline]
Abbreviations
BMI: Body Mass Index |
ROM: range of motion |
3D: three-dimensional |
Edited by G Eysenbach; submitted 04.12.14; peer-reviewed by G Kurillo, G Ferriero; comments to author 15.02.15; revised version received 24.03.15; accepted 26.04.15; published 20.05.15
Copyright©Cristina Roldan-Jimenez, Antonio Cuesta-Vargas, Paul Bennett. Originally published in JMIR Rehabilitation and Assistive Technology (http://rehab.jmir.org), 20.05.2015.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Rehabilitation and Assistive Technology, is properly cited. The complete bibliographic information, a link to the original publication on http://rehab.jmir.org/, as well as this copyright and license information must be included.